Medtronic Covidien Valleylab Force TRIAD
Medtronic Covidien Valleylab Force TRIAD Features
The Covidien Force Triad energy platform is a full-featured
electrosurgical system that provides electrosurgical cutting and
coagulation, bipolar functionality, and vessel sealing in a
single generator. This Electrosurgical Unit is intended for open
and laparoscopic surgical procedures and includes; TissueFect,
ValleyLab, and LigaSure fusion technologies. It features a
user-friendly interface with three touchscreens that allow the
user to control the system functions.
TissueFect sensing technology to precisely
manage energy delivery.
Enhanced LigaSure tissue fusion technology
that fuses vessels up to 7mm, lymphatics, pulmonary vasculature
and tissue bundles.
alleyLab mode that provides a unique
combination of monopolar hemostatis and dissection.
Automatic instrument recognition.
Compatible with conventional electrosurgical
instruments and all current and new LigaSure instruments.
Medtronic Covidien Valleylab Force Triad Specifications:
General
Output Configuration: Isolation output.
Cooling: Natural convection and fan.
Display: Three LCD touchscreens.
Connector ports: LED illuminated Smart
connector readers
Force Triad energy
platform cart (FT900) Universal.
Mounting cart
(UC8009), and/or the UC8010 Over-shelf.
Operating room boom
systems.
Any Stable, flat
surface such as a table or cart top.
Dimensions and Weight
Height: 25.5 cm (10 in.)
Width: 45.8 cm (18 in.)
Depth: 50.8 cm (20 in.)
Weight: 13.6 kg (30 lbs)
Activation Tone
The audio levels stated below are for
activation tones (cut, Valleylab, coag, bipolar, and LigaSure
Modes) and alarm tones (REM and system alarms) at a distance of
one meter.
Volume and Frequency
Volume (adjustable): 45 to 65 dBA
Frequency Cut: -660 Hz
Valleylab: 800 Hz.
Coag: 940 Hz.
Bipolar: 940 Hz.
LigaSure: 440 Hz.
Duration: Continuous while the system is
activated.
Alarm Tone
Volume (adjustable): >65 dBA.
Frequency: -REM: 660 Hz.
Regrassp: Two tones:
High= 985 Hz, Low = 780 Hz.
Seal Complete: 985
Hz.
Error/System Alert:
Beep tone = 1400 Hz.
Duration:
REM: Two 1/2 second tones separated by 1/2 second for each REM
event
Reactivate/Regrassp: Four 175 msec tones-high, low, high–low
separated by 1/2 second.
Check Instrument: Six 175 msec tones – high, low, high low, high,
low -Seal Complete: Two 175 msec tones separated by 175 msec for
each Seal Complete event.
Error/System Alert: Two 1/2 second tones separated by 1/2 second
for each Error/ System Alert event.
REM Contact Quality Monitor
Interrogation frequency: 80 kHz ± 10 kHz
Interrogation current: < 100μA
Interrogation voltage: < 12V RMS
Acceptable Resistance Range
REM resistance measurements are ± 10% during
RF activation and ± 5% when RF output is not activated.
REM patient return electrode: 5 to 135 ohms or
up to a 40% increase in the initial measured contact resistance
(which ever is less).
If the measured resistance is outside the
acceptable range(s) noted above, a REM fault condition
occurs.
REM Alarm Activation
REM patient return electrode: When the
measured resistance exceeds the standard range of safe resistance
(below 5 ohms or above 135 ohms) or when the initial measured
contact resistance increases by 40% (whichever is less), the REM
alarm indicator enlarges and flashes red and yellow, a tone
sounds twice, and RF output is disabled. The indicator remains
illuminated red and yellow until you correct the condition
causing the alarm. Then, the indicator illuminates green and RF
output is enabled.
Autobipolar
The ForceTriad energy platform is equipped
with an autobipolar feature that allows for automatic activation
of bipolar energy.
Note: The autobipolar electrode function
requires the use of the Valleylab E0018 bipolar Instrument
cord.
Autobipolar Specificaitons:
Interrogation
frequency: 80 kHz ± 10 kHz.
Interrogation current:
< 100 μA.
Activation impedance:
20 Ω to 500.
Deactivation
impedance: User selectable: 1,500 Ω 1,800 Ω or 2,200 Ω.
Measurement accuracy:
± 14% of Full Scale activation impedance while keying active
above 10 watts.
+ 50%, -14% of full
scale activation impedance while keying active at 10 watts or
below.
± 5% of Full Scale
activation impedance while keying inactive.
Keying delay: User
selectable in 500 msec increments from 0 sec to 2.5 sec
Duty Cycle
Under maximum power settings and rated load
conditions, the ForceTriad energy platform is capable of
operating a duty cycle of 25% defined as 10 seconds active and 30
seconds inactive, in any mode for a period of 4 hours.
Caution
Use of duty cycles
greater than 25% (10 seconds active followed by 30 seconds
inactive) will increase the risk that heat build-up under a
return electrode may be high enough toinjure the patient. Do not
continuously activate for longer than one minute.
Low Frequency (50/60 Hz) Leakage Current
Enclosure source current, ground open