Siemens Mobilett XP Portable X-Ray Machine
Siemens Mobilett XP Portable X-Ray Machine Features
The Siemens Mobilett XP features an integrated cable design –
eliminating loose cables that can get tangled or caught on
objects during transport. The integrated cable design on the
Siemens Mobilett XP also enables easy arm positioning and makes
the system easy to clean.
The lightweight MOBILETT XP systems are equipped with
polyurethane tires on a short wheelbase to help overcome any
floor resistance. A convenient handbrake stops the system safely.
More to that, the compact design allows transportation in
elevators and other areas where space might be limited, for
example in an ICU.
The sleek design always allows good visibility to the front.
The tube head of the system is mounted on a double articulated
and fully counterbalanced X-ray arm with integrated cables for
maximum cleanability. Its industry-leading horizontal extension
of up to 128 cm and the freely rotatable head make MOBILETT XP
easy to position in virtually any projection, with the patient
recumbent, sitting or any position in between. This is of special
importance when you have to deal with immobile and critically ill
patients.
The self-calibrating high imaging output, up
to 30 kW and 360 mAs (for 230 V line operation), enables short
exposure times down to 1 ms – virtually eliminating motion
blurring.
Remarkably sharp images, made possible by the
short exposure times, enhance its utility in chest imaging and
neonatal units
The Siemens Mobilett XP is the optimal choice
for semi-sterile environments such as ICU, CCU, neonatal and
pediatric departments, or the wards
The X-ray examination is easy to complete with
the MOBILETT XP
Siemens Mobilett XP Portable X-Ray Specifications
Dimensions
Height: 61 in (155 cm)
Width: 23 in (57.5 cm)
Length: 39 in (98.7 cm)
Weight: Aprox. 206 kg
Tube Positioning
Max. focal spot height 2135 mm (84.1 inch)
Min. focal spot height 478 mm (18.8 inch)
Max. horizontal ext. 1285 mm (50.6 inch)
Min. horizontal ext. 282 mm (11.1 inch)
Output
max. 133 kV 225 mA
max. 450 mA 54 kV
max. 30 kW – 99 kV – 303 mA – 10 ms
max. 25 kW – 99 KV – 252 mA – 100 ms (IEC
60601-2-7)
min. 0.32 mAs
Line Voltage
100–130 V ±10 %, 50/60 Hz, impulse current
11A,
continuous current 0.6 A
200–240 V ±10 %, 50/60 Hz, impulse current
19A,
continuous current 0.5 A
Apparent resistance of supply AC line
100–130 V < 0.6 Ohm
200–240 V < 2.5 Ohm
Environment
Opertion
Permissible ambient
temperature: +10 °C – +40 °C
Permissible relative
humidity: 30 % – 75 %
Permissible air
pressure range: 800 hPa
Transport/Storage
Permissible ambient
temperature: -20 °C – +70 °C
Permissible relative
humidity: 10 % –100 %*
Permissible air
pressure range: 1060 hPa
Stabilization
Microprocessor-controlled adjustment and
regulation during exposure
X-Ray Tube
Siemens P135/30 R with nominal focal spot
value 0.8 (acc. to IEC 60336)
Anode Rotation
9,000 r.p.m., 142 Hz (anode angle 15°)
Anode Heat Storage Capacity
90,000 J = 122,000 HU
Tube Housing Heat Storage Capacity
800,000 J = 1,100,000 HU
Inherent Filtration
Collimator 2.0 mm Al (at 70 kVp) min.
Tube housing assembly 1.9 mm Al (at 70 kVp)
min.
Total min. 3.9 mm Al according to IEC
60522
Collimator
Manually adjustable multileaf collimator,
rotatable ± 90°
Collimator light and light field intensity
Halogen lamp min. 180 lux at 1 m SID, without
DAP equipment
Measurement of Leakage radiation
133 kVp/10 mAs, 200 exposures/h