Product Details
Product Introduction
TESTech Limit Oxygen Index Tester is designed according to
ISO4589.1-2006, it applies to analysis the minimum oxygen
concentration needed to maintain combustion of the sample when
the sample is under under the test conditions specified in oxygen
and nitrogen mixed gas;
This tester is installed with England imported Oxygen sensor, and
automatic control the Oxygen and nitrogen flow with mass flow
control meter. It is very easily and quickly to operate; Used for
evaluating homogeneity of solidmaterials, laminates, foam, film
and thin film in the specified conditions of flammability.
Model:FTech-ISO4589
Standard reference:
ISO4589.-2006
Technical Parameter:
1, Oxygen concentration measurement range: 0-100%;Accuracy of ±
2, Oxygen concentration 0.1%
2, Oxygen Mass flow meter range: 0~15.0L/min
3, Nitrogen Mass flow meter range: 0~15.0L/min
4, Response time: <10s
4,Oxygen concentration measurement accuracy: ± 0.4%
5, Combustion tube top airflow 90mm ± 10MM/s
6,Combustion air flow in cylinder 6, 40mm/s ± 2mm/s,
concentration control accuracy ± 0.1%
7,Measurement time, accuracy ± 0.5S
8,Portable igniters, Jet flame down 16± 4mm adjustable;
9,Double flow meter and pressure gauge devices: precision: 1%
Features:
1,England Imported ParaMagnetic oxygen sensor, accuracy is ± 0.1%
; without working life limit, it is definitely difference with
electrochemistry oxygen sensor. (It is difference with
electrochemistry sensor, no need to change the sensor every
year.)
2,Equipped with heat-resistant quartz glass tube, export ID
40mm;
3,Equipped with supporting and not supporting sample holder
4,Mass flow controllers, automatic regulating oxygen and nitrogen
gas flowaccording to the set concentration;
5,Patent design of oxygen and nitrogen mixed unit to make sure
the gas is mixed entirely
6,The expected oxygen concentration is set on screen, and PLC
control system; No need manually operation;
7, LCD screen show test data such as flow rate, time, results in
real time and oxygen concentration etc .
Control System:
1, PLC with touch screen system to make the test high
automation;
2, After setting the oxygen concentrations, the system
automatically adjustingthe oxygen concentration;
3, Key electrical components adopt imported brands, accurate,
reliable, anddurable;
4, Equipped with a protection circuit;
Smoke Density System(Optional)
1, Optical absorption range: 0-100% can be measured continuously.
Smoke density resolution 0.1%
2, Measurement of smoke density is accuracy: ± 3%;
3, Optical glass is easy to be cleaned, avoid to the negative of
dust particles;
4, Light resource is 6V instrument special purpose incandescent
lamps;
5, System is equipped with calibration system for light
transmittance;