Description:
This tester is used to test the acidity of the gas that released
during combustion from cable or optical cable components
material; The equipment is designed and produced on the basis IEC
60754.The cable material is processed by prius tube furnace, and
then the pH value and conductivity of the gas that generated in
combustion process are measured,so that the cable material of
acidic,as well as concentration of halogen acid gas volume, PH,
conductivity, etc, can be determined.
Model: FTech-IEC60754
Standards:
• IEC 60754-1 Ed. 2.0 b:2011
• IEC 60754-2 Ed. 1.0 b:2011
Main Features
1, The tester designed with optimized structure, logical layout
and easy to operate.
2, Al the fixtures are made of stainless steel.
3, The effective length of the tube-type furnace is 500~600mm,
inner diameter is 40~60mm, with adjustable electric systems.
4, Equipped with imported thermocouple from United States.
5, The size of the combustion boat is length 45~100mm, width
12~30mm, depth 5~10mm.
6, The electrical conductivity measuring range is
0.01uS/mm~100uS/mm.
7, The control system using an electronic temperature control,
intelligent control tube furnace temperature.
8, Equipped with two gas collection bottles.
9, The size of the sample chamber is 1 liter.
10, Equipped with a PH value and conductivity detector, with an
electronic display and an electrode.
11, Air flow meter is equipped with silicon dioxide fire pipeline
to prevent corrosion by the gas.
12, Activated carbon (air filter): filter air through a
filter.
13, The silicone is used for drying air source.
14, Pyrex pipeline: Users can easily insert sample into the Prius
tube to protect the pipeline from damage.
15, Combustion boat moving system: combustion boat moves
automatic to achieve more intelligent test.
16, The mode switching: mode switch (two test methods to choose
from, IEC 60754-1 & IEC 60754-2).
17, The temperature alarm system: control the furnace temperature
against overheating.
18, Timer: IEC 60754-1 or IEC 60754-2 time measurement
device.
19, Drawer:used to store parts of the box (such as conductivity
meters, pH meter, etc.)
20, Temperature indicator: shows the furnace core
temperature.
21, Permanent magnetic stirrer: The agitator is used to stir the
combustion gases in distilled water.
22, Conductivity meter:An instrument which used to measure the
combustion gas conductivity.