| MOQ | 1 PC |
| 1.5kgs | G.W. |
| Range of flow rate setting |
0.01-1800ml/h, with resolution 0.01ml/h |
| Flow rate accuracy (Essential Performance) | ±3% |
| VTBI range |
0.01~9999.99ml, with resolution 0.01ml |
| Infusion volume accuracy (Essential Performance) | ±3% |
| Purge speed | 1ml/h~1800ml/h;±20% |
|
Occlusion
alarm(pressure)
(Essential Performance)
|
High: 100kPa ± 30kPa middle 60 kPa ± 20kPa Low: 40kPa±20kPa |
| Maximum infusion pressure | >160kPa |
| KVO flow rate |
<10ml/h; KVO=the set infusion flow rate when flow rate <1ml/h. |
| Recover time after the recoverable alarm sound is cleared. | 1min 50 s~ 2min |
| Time for pause over time alarm. | 1min 50 s~ 2min |
| High accuracy |
The accuracy for both infusion flow rate and volume
are kept within
3% when
the
UNM
recommended infusion set is used
|
| High flow rate |
Infusion flow rate can be adjusted from 0.1ml/h to
1800ml/h in a
continuous manner, which
makes
UNM20 capable of meeting various flow rate requirements in different infusion cases. |
| Small size | Just 6.8cm tall and 1.5kg in weight, UNM20 is not only very small in dimension but also very light. |
| Stackable | UNM20 pump is stackable. It can also be stacked with UNM31 syringe pump for operation. The miniature design of UNM20 is a room saver for the wards where space is very limited. It can also be inserted onto an MX infusion work station as an infusion unit. |
| Easy operation | Operator can use the touch screen onUNM20 to set parameters, which will still function with gloves on. A key pad is also available to ensure usability in different usage scenarios. |
| Powered free-flow clamp | UNM20 has a powered free-flow clamp that saves several steps in the infusion set installation process. To complete the installation process, the operator only needs to straighten the infusion set with both hands, clamp it at two ends, and close the pump door. |
|
Upstream occlusion
alarm |
UNM20
has upstream occlusion alarm in addition
to
downstream occlusion alarm. When infusion bag is
running out or the free-flow
clamp
is n ot turned on by mistake, the embedded internal
pressure sensor will detect these problems
automatically and an alarm will be initiated
accordingly.
|
|
Fast
installation |
Patented Quick Mount system, which requires only one click to complete the pump installation. |
|
External power
source |
An external power adapter is used, which not only removes the safety concerns of using an internal power source but also makes the device lighter, safer, and more portable. |
|
High battery
capacity |
The rechargeable internal high-capacity Lithium battery can support normal operation for 7 hours, which is conveniently helpful during patient transport or power outage. |
| Highly secure STM32 microcontroller | dual-CPU architecture design |
|
No false
alarm in air-in-line
detection |
Based on ultrasonic technology and with the help from a unique algorithm, the air-in-line detection is accurate and reliable, which eliminates false alarms. |