MOQ | 1 unit |
Port | Frihamnen, Sweden |
Packaging | boxes and pallets |
Lead Time | 5 - 7 days |
coronavirus | treatment |
Parameters for icu ventilator
tidal volume, ventilation volume, IPPV rate, SIMV rate,
total respiratory rate, I/E, peak pressure of airway, average pressure,
pressure - time waveform, flow rate - time waveform, inspiratory trigger pressure,
Inspiratory platform (autonomous inspiratory rate,
lung compliance, airway resistance, flow -volume loop, pressure-volume loop)
Monitoring of oxygen concentration 21%~100%
the feature of the icu ventilator
Security Alarm System of icu ventilator
Oxygen Concentration alarm upper limit setting range 21%~100%
low limit setting range 10%~80%
Airway pressure alarm upper limit setting range 0.3~6.0 kPa
low limit setting range 0.2~5.0 kPa
Per-minute ventilation volume alarm upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm it will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarm it will give sound and light alarm if there is
no tidal volume input for 15 seconds
Power alarm
Power AC 220V±10% 50Hz±1Hz
UPS, Storage battery
Mechanical Arms
(Air compressor)
Items within brackets are optional
the specification of the icu ventilator
Display mode High-definition 10.4TFT color LCD screen display
Ventilation mode IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH
Ventilation parameters Tidal volume50~1500 ml (20~1500 ml)
Rate 1~99 bpm
SIMV rate 1~20 bpm
I:E 2:1~1:8 (4:1~1:8)
Inspiratory trigger pressure: -1.0~2.0Kpa
0~2.0Kpa
Pressure range 1.0~6.0Kpa
SIGH 1.5 times the inspiratory time(70-150 adjustable)
Inspiratory platform 0~50%