Steris V-PRO maX 2
Steris V-PRO maX 2 Features
The Steris V-Pro maX 2 is a low temperature sterilization system
that uses VAPROX HC Sterilant. It properly prepares (cleans,
rinses, and dries) medical devices in healthcare facilities.
V-Pro max 2 come with preprogrammed sterilization cycles that
operate at low pressures and temperatures for processing medical
devices without leaving toxic residues.
Large, easy to use touch screen control
panel.
Proprietary hydrogen peroxide based
sterilant.
Process monitoring and cycle
documentation.
Specially designed conditioning phase that
aids in the removal of residual moisture.
Easy installation.
Steris V-PRO maX 2 Specifications
General
Chamber Volume (ft3): 4.3 (136L)
Cycle Time: 16 Fast non Lumen; 28 Non Lumen;
52 Lumen; 35 Flexible
Moisture Check: Automatically included in
cycle time
Fast BI: 20 minutes
Cycles per Sterilant Container: 15 Cycles (or
30 for Fast Non Lumen Cycle)
H2O2 Concentration: 59%
Plasma: No
Compatible Materials: 40
Configurations: Single Door or Double Door
Laparoscopy Trays per Shift: 68
Connect to Tracking Systems: Yes
Device Configurations
Single-channel
Stainless Steel: ≥
0.77mm x ≥ 500mm
Flexible: ≥ 1mm x
≥1050mm (Max 2 per load)
Dual-channel
Stainless Steel: ≥
0.77mm x ≥ 527mm
Flexible: ≥ 1mm x ≥
990/850mm (Max 2 per load)
Triple-channel: Stainless Steel ≥1.2mm x ≥
275mm; ≥ 1.8mm x ≥ 310mm; ≥ 2.8mm x 317mm
Flexible Plus Load: 1 Scope (single or dual)
and non-lumened load for a total of 24 lbs.
Max Lumens/Load: 20