Detector channels: | 192*64 LCD display | Color | white |
●Advanced percentage analysis principle overcome the interfere with Jaundice,
Hemolysis, and Chyle. Precision testing on Abnormal samples
●Mini- reagent consumption up to 20ul, reagent open system
●Special location setting system of testing cups,loading bracket, precision operating
●Special measuring cell diaphragm system,
stop other light interfere
●Perform all routine tests such as PT, APTT, TT Fibrinogen and clotting factors Provide 2 kind of analytical method of FIB
●Can set more than 5 point of standard curve, modify and saving parameters, variety item can be added
●Online startup and manual startup system
●Incubate management system with audible signal hint
●Built-in thermal printer with reference value range
●Large memory for 10000 results
●Detector channels: 1
●Position for Reagents: 2
●Position for Sample: 5
●Light source: 470nm long-life cold light source
●Sample consumption: 20-40uL
●Mini-reagent consumption: 20uL
●Homoiothermy controlling: 37±0.3 degrees centigrade
●Repeatability: CV≤2%
●Linear error: r>0.98
●Measurement startup: online adding, auto timing, manual startup system
●Memory: 10000 results
●Counting function: can be converted on S, %, PTR, INR, mg/dL
●Printer: built in thermal printer
●Display: 192*64 LCD display
●Working Condition:Temperature:10-30 degrees centigrage,
Relative Humidity:20%-80%
●Voltage:AC 90~264V,auto adapt according to the input voltage
●Size: 280mm(L)*300mm(W)*160mm(H)