Port | Shenzhen bay |
Packaging | Paperboard case. 70x40x40 cm. |
Lead Time | 5 |
M-100 | Yellow and white |
*Different language versions are available
It is especially suitable for the process that requires frequent detection of the corresponding substance concentration in the fermentation industry. M-100 Automated biosensor analyzer is very fast, stable and cost-effective, many customers have bought and used them for more than 3 years, it is the first choice for mammalian cell or hybridoma cell culture process development.
This device is available for detecting the concentration of the test sample by use of the specially designed bio-oxidase membrane electrochemical sensor.
M-100 Automated biosensor analyzer could automatically collect samples and import them into the test area. The test substance contained in the sample will undergo an enzymatic hydrolysis reaction under the catalysis of an immobilized bio-oxidase to generate hydrogen peroxide. The concentration of the test substance is calculated by the hydrogen peroxide concentration that is generated in the detection of hydrogen peroxide electrode.
The device can calibrate a standard item with a known concentration, and the voltage value of the standard item is a measure of the concentration of the test substance. The unknown concentration can be obtained through the comparison with the voltage signal of the standard item. The sensor electrodes will be automatically cleaned by the system buffer solution after each measurement is completed, and the next test can be performed after cleaning.
Test principle | Enzyme electrode method |
Electrode | Rod electrode |
Measuring range |
Glucose(G1001): 0.05~1
g/L
Glucose(C1001): 0.3~9 g/L Lactate: 0.03~2 g/L Glutamate: 0.03~1 g/L Lysine: 0.03~1 g/L Methanol: 0.03~0.5 g/L Ethanol: 0.03~1 g/L Xylose:0.03~1 g/L Glycerin:0.03~1 g/L Glutamine:0.03~1 g/L |
Resolution | 0.01g/L |
Precision (CV%) | <2% |
Enzyme membrane life |
Glucose:
≤30
days, Lactate:
≤30
days, Glutamate:
≤30
days,
Lysine: ≤30 days, Ethanol / methanol: ≤10 days, Xylose: ≤15days, Glycerin: ≤15days, Glutamine:≤15 days |
Ambient temperature | 10~35°C |
Relative humidity | ≤85% (no condensation water) |
External dimension | 480×320×210 mm |
Power supply | 110V or 220V, 50Hz, 60Hz, separate three-hole socket with good grounding |
Power | <100VA |
Weight | 9kg |
Screen | 8-inch color touch screen |
*Automatic injection and detection
Test items: Glucose
Detection range: Glucose: 0-9g/L(0-50mmol/L)
Test items: Lysine
1. Pre-treatment: taking advantage of the property that lysine
hydrochloride is easily soluble in water (40g/ml,35℃), 1.00g of
98% commercially available feed grade L-lysine hydrochloride
sample was taken and fully dissolved, with a constant volume of
100mL, and then tested on the machine
2. Results
Brand | 1 | 2 | 3 | 4 |
Lysine analyzer, % | 98.88 | 98.92 | 98.85 | 98.99 |
HPLC,% | 98.63 | 98.89 | 98.52 | 98.68 |
Perchloric acid titration, % | 98.96 | 98.82 | 99.11 | 98.71 |
Building 14, Xusheng Ronghe Valley, No.9 Furong Road, Songgang, Baoan District, Shenzhen