Mask Bacterial Filtration Efficiency (BFE) Detector

MOQ1 Unit
PortGuangzhou Huangpu
PackagingWooden box
Lead Time20 working day

Quick Details (View All)

Colorwhite

Product Details

  • Application

It is used to test the percentage of materials containing bacteria suspended particles at the specified flow rate. The dual gas path simultaneous comparison sampling method is used to improve the accuracy of sampling, which is suitable for the performance testing of bacterial filtration efficiency of medical surgical masks by metrological inspection departments, scientific research institutes, medical mask manufacturers and other relevant departments.

  • Standards

YY0469-2004YY/T 0969-2013ASTMF2100ASTMF2101EN14683

  • Specifications

Key Specification

Specification Range

Resolution

Accuracy

A Route Sampling Flow

28.3L/min

0.1L/min

Within ± 2.5%

B Route Sampling Flow

28.3L/min

0.1L/min

Within ±2.5%

Spray Flow

(8~10)L/min

0.1L/min

Within ±2.5%

Peristaltic Pump Flow

(0.006~3.0)mL/min

0.001ml/min

Within ±2.5%

Front Pressure of A Route Sampling Flowmeter

(-20~0)kPa

0.01kPa

Within ±2.5%

Front Pressure of B Route Sampling Flowmeter

(-20~0)kPa

0.01kPa

Within ±2.5%

Front Pressure of Spray Flowmeter

(0~300)kPa

0.1kPa

Within ±2.5%

Ambient Temperature

(-40~99)

0.1

Within ±2.5%

Negative Pressure of the Aerosol Chamber

(-90~-120)Pa

0.1Pa

Within ±2.0%

Negative Pressure of the Cabinet

-50~-200pa

Data Storage Ability

>100000 sets

Vortex Mixer Test Tube Specification and Quantity

Φ16×150mm test tube8pcs

High Efficiency Particulate Air Filter Properties

Filter Efficiency≥99.99% for particles> 0.3μm

Median diameter of aerosol generator mass

Average diameter: (3.0±0.3)μm; Geometric Standard Deviation≤1.5

Double Routes 6-lever Andersen Sampling apparatus

Level1>7μm; Level 2: (4.7~7)μm;

Level 3: (3.3~4.7)μm; Level 4: (2.1~3.3)μm;

Level 5: (1.1~2.1)μm; Level 6: (0.6~1.1)μm

Size of Aerosol Chamber

600×85×3mm (Length×Diameter×Thickness)

Total number of positive quality control sampler particles

(2200±500)cfu

Negative Chamber Flow Speed

≥5m3/min

Size of Negative Pressure Chamber Door(W×D)

1000×730mm

Size of Main Machine(W×D×H)

1180×650×1300mm

Size of Supporting Frame(W×D×H)

1180×650×600mm, Height within 10cm is adjustable

Noise of the Tester

<65dB(A)

Weight

150kg

Power Supply

AC220V±10%, 50Hz

Power Consumption

<1500W

    • Features

1. Negative pressure experiment system to ensure the safety of operators;

2. Negative pressure cabinet with built-in peristaltic pump and A and B two-way six-stage Andersen;

3. The flow rate of the peristaltic pump can be set;

4. Dedicated microbial aerosol generator can set the volume of bacterial liquid spray flow, and the atomization effect is good;

5. Embedded high-speed industrial microcomputer control;

6.10.4-inch industrial-grade high-brightness color touch display;

7, USB interface, support U disk data transfer;

8. Built-in high-brightness lighting in the cabinet;

9. Built-in leakage protection switch to protect operator safety;

10. Insulation and flame retardant between inner and outer layers;

11. The front switch glass door is convenient for experimenters to observe and operate;

12. Detachable stand, the height of stand is adjustable;

13. Support and move dual-purpose casters.